Sarns 8000 Manual

-
SarnsSarns 8000 manual 2020HeartSarns 8000 manualManual
510(k)|DeNovo|Registration & Listing|Adverse Events|Recalls|PMA|HDE|Classification|Standards
CFR Title 21|Radiation-Emitting Products|X-Ray Assembler|Medsun Reports|CLIA|TPLC


Sarns 8000 Manual Online

Terumo Sarns 8000; Service Manuals; saad. Sat Jul 24 2021 Service Manuals. I need operating & service manual of sarns 8000 heart lung machine. Sarns8000Manual 1/5 PDF Drive - Search and download PDF files for free. Sarns 8000 Manual Sarns 8000 Manual Recognizing the pretentiousness ways to acquire this ebook Sarns 8000 Manual is additionally useful. You have remained in right site to begin getting this info. Acquire the Sarns 8000 Manual join that we offer here and check out the link.

Sarns 8000 Manual Free

New SearchBack to Search Results
Class 2 Device Recall Sams Modular Perfusion System 8000
Date Initiated by FirmSeptember 25, 2012
Date PostedNovember 13, 2012
Recall Status1Terminated 3 on July 11, 2013
Z-0281-2013
Recall Event ID62733
K915183
Console, heart-lung machine, cardiopulmonary bypass - Product CodeDTQ
ProductPerfusion System 8000, base 5 pump, 115V safety monitor included
The Sarns' Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Code Informationcatalog number: 16401 serial numbers: 1001-1243, 1247-1648, 1650-1821, and 1824-3003.
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corporation
6200 Jackson Road
Ann Arbor MI 48103-9586
For Additional Information Contact
734-741-6173
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular Perfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure alarms. If the malfunction of the arterial or cardio
Device Design
The firm, Terumo CVS, sent an 'URGENT MEDICAL DEVICE RECALL' notice/Safety Advisory Alert dated September 24, 2012 and an addendum to the Operator's Manual to alert all affected Customers. The notice described the product, problem and actions to be taken. The customers were instructed to review this medical device recall notice; assure that all users are aware to this notice; place the addendum in the Operator's Manual just before page 4.1 and complete and return the attached Customer Response Form via Fax to: 1-734-741-6149 or email to:www.termuo-cvs.com. If you have any questions or concerns contact Terumo CVS Customer Service at 1-800-521-1818; Monday - Friday, 8 AM - 6 PM ET.
1097
Worldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY; and countries of: ARGENTINA, AUSTRALIA, BANGLADESH, BRAZIL, CALI COLOMBIA, CALLO PERU, GERMANY, GUAYQUIL ECUADOR, HONG KONG, INDIA, INDONESIA, JAPAN, LEBANON, MADRID SPAIN, MEXICO, NEW ZEALAND, SANTIAGO, SANTIAGO CHILE, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SRI LANKA, TAIWAN, THAILAND, and UNITED ARAB EMIRATES (UAE).
TPLC Device Report
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 ยง7.55.
510(K) Database510(K)s with Product Code = DTQ and Original Applicant = 3M HEALTH CARE, SARNS